MétaCan
Menu
Back to cohort
Record W2775082052 · doi:10.1177/0194599817742356

Analysis of Reported Balloon Malfunctions and Proposed Rescue Strategy for Malfunction during Airway Dilation

2017· article· en· W2775082052 on OpenAlexaff
E. Brandon Strong, Derrick R. Randall, Daniel J. Cates, Peter C. Belafsky

Bibliographic record

VenueOtolaryngology · 2017
Typearticle
Languageen
FieldMedicine
TopicTracheal and airway disorders
Canadian institutionsUniversity of Calgary
FundersUniversity of California, Davis
KeywordsDilatorMedicineBalloonAirwaySurgeryBalloon dilationRadiology

Abstract

fetched live from OpenAlex

Objective The rate of balloon dilator failure is unknown, and a rescue strategy for device malfunction has not been established. The purposes of this investigation were to determine the approximate number of balloon failures in the gastrointestinal tract and airway, evaluate the parameters required to rupture balloon dilators, and develop a rescue strategy to efficiently reestablish airway patency. Study Design Retrospective cohort and basic medical research. Setting Academic tertiary care medical center. Subjects and Methods The Manufacturer and User Facility Device Experience database was queried for adverse events associated with tracheal and esophageal dilators between January 1, 2014, and January 1, 2017. A bench-top model of airway stenosis was developed, and optimal conditions for the safe removal of a malfunctioning dilator were assessed (2, 4, 6 atm). Results There were 420 reported balloon malfunctions, including 104 cases with deflation/removal issues. The bench-top model determined that device rupture allowing for immediate removal occurs with needle puncture at balloon pressures ≥8 atm. Balloons inflated to 6 atm required a median of 17.5 seconds (range, 0-55.3) for removal, in comparison with 30.2 seconds (range, 7.1-87.5) at 2 atm ( P > .05). Conclusion Balloon dilator malfunction is a significant problem that practitioners must be prepared for. Pressure ≥8 atm (~33% overinflation) is required to consistently cause complete balloon dilator rupture via needle puncture. While counterintuitive, increasing the inflation pressure of a malfunctioning balloon (8-10 atm) may expedite rupture and safe removal. A rescue strategy for balloon malfunction is proposed.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.072
Threshold uncertainty score0.376

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.031
GPT teacher head0.300
Teacher spread0.269 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations8
Published2017
Admission routes1