Performance of an enhanced afterloader with electromagnetic tracking capabilities for channel reconstruction and error detection

2021· article· en· W3158902066 on OpenAlexaffabout

Bibliographic record

VenueMedical Physics · 2021
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsUniversité LavalCentre hospitalier universitaire de Québec
Fundersnot available
KeywordsTracking (education)Channel (broadcasting)Identification (biology)Tracking errorMeasurement uncertaintyError detection and correctionCatheter

Abstract

fetched live from OpenAlex

PURPOSE: To assess catheter reconstruction and error detection performance of an afterloader (Elekta Brachytherapy, Veenendaal, The Netherlands) equipped with electromagnetic (EM) tracking capabilities. MATERIALS/METHODS: The Flexitron research unit used was equipped with a special check cable integrating an EM sensor (NDI Aurora V3) that enables tracking and reconstruction capability. The reconstructions of a 24-cm long catheter were performed using two methods: continuous fixed-speed check cable backward stepping (at 1, 2.5, 5, 10, 25 and 50 cm/s) and stepping through each dwell position every 1 mm. The ability of the system to differentiate between two closely located (parallel) catheters was investigated by connecting catheters to the afterloader and moving it from its axis with an increment of 1 mm. A robotic arm (Meca500, Mecademic, Montreal) with an accuracy of 0.01 mm was used to move the catheter between each reconstruction. Reconstructions were obtained with a locally weighted scatterplot smoothing algorithm. To quantify the reconstruction accuracy, distances between two catheters were computed along the reconstruction track with a 5 mm step. The reconstructions of curve catheter paths were assessed through parallel and perpendicular phantom configuration to the EM field generator. Indexer length and lateral errors were simulated and a ROC analysis was made. RESULTS: Using a 50 cm/s check cable speed does not allow for accurate reconstructions. A slower check cable speed results in better reconstruction performance and smaller standard deviations. At 1 cm/s, a catheter can be shifted laterally down to 1 mm and all paths can be uniquely identified. The optimum operating distance from the field generator (50 to 300 mm) resulted in a lower absolute mean deviation from the expected value (0.2 ± 0.1 mm) versus being positioned on the edge of the electromagnetic sensitive detection volume (0.6 ±0.3 mm). The reconstructions of curved catheters with a check cable speed under 5 cm/s gave a 0.8 mm ±0.3 mm error, or better. All indexer and lateral shifts of 1 mm were detected with a check cable speed of 2.5 cm/s or lower. CONCLUSIONS: The EM-equipped Flexitron afterloader is able to track and reconstruct catheters with high accuracy. A speed under 5 cm/s is recommended for straight and curved catheter reconstructions. It allows catheter identification down to 1 mm inter-catheter distance shift. The check cable can also be used to detect common shift errors.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.002
metaresearch head score (Gemma)0.006
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.002
Threshold uncertainty score0.011

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0020.006
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0010.001
Open science0.0010.001
Research integrity0.0010.000
Insufficient payload (model declined to judge)0.0020.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.007
GPT teacher head0.245
Teacher spread0.238 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations12
Published2021
Admission routes2
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207