The Accuracy of Pulse Oxygen Saturation, Heart Rate, Blood Pressure, and Respiratory Rate Raised by a Contactless Telehealth Portal: Validation Study

2024· article· en· W4393956330 on OpenAlexvenueno aff
Julian Gerald Dcruz, Paichang Yeh

Bibliographic record

VenueJMIR Formative Research · 2024
Typearticle
Languageen
FieldEngineering
TopicNon-Invasive Vital Sign Monitoring
Canadian institutionsnot available
Fundersnot available
KeywordsTelehealthHeart rateOxygen saturationRespiratory ratePulse rateBlood pressureMedicineRespiratory systemCardiologyPulse (music)Saturation (graph theory)Materials scienceOxygenBiomedical engineeringInternal medicineComputer sciencePhysicsMathematicsTelemedicineTelecommunicationsPolitical science

Abstract

fetched live from OpenAlex

Background The traditional measurement of heart rate (HR), oxygen saturation (SpO2), blood pressure (BP), and respiratory rate (RR) via physical examination can be challenging, and the recent pandemic has accelerated trends toward telehealth and remote monitoring. Instead of going to the physician to check these vital signs, measuring them at home would be more convenient. Vital sign monitors, also known as physiological parameter monitors, are electronic devices that measure and display biological information about patients under constant monitoring. Objective The purpose of this study was to validate the accuracy of the pulse SpO2, HR, BP, and RR raised by Docsun Telehealth Portal by comparing it with approved medical devices. Methods This is a noninvasive, self-check, system-based study conducted to validate the detection of vital signs (SpO2, HR, BP, and RR) raised by Docsun Telehealth Portal. The input for software processing involves facial screening without any accessories on the face, scanning directly through the software application portal. The participant’s facial features are detected and screened for the extraction of necessary readings. Results For the validation of HR, SpO2, BP, and RR measurements, the main outcomes were the mean of the absolute difference between the respective investigational devices and the reference values as well as the absolute percentage difference between the respective investigational devices and the reference values. If the HR was within ±10% of the reference standard or 5 beats per minute, it was considered acceptable for clinical purposes. The average absolute difference between the Docsun Telehealth Portal and the reference values was 1.41 (SD 1.14) beats per minute. The mean absolute percentage difference was 1.69% (SD 1.37). Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for HR measurements. If the RR was within ±10% of the reference standard or 3 breaths per minute, it was considered acceptable for clinical purposes. The average absolute difference between the Docsun Telehealth Portal and the reference values was 0.86 breaths per minute. The mean absolute percentage difference was 4.72%. Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for RR measurements. SpO2 levels were considered acceptable if the average absolute difference between the Docsun Telehealth Portal and the reference values was ±3%. The mean absolute percentage difference was 0.59%. Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for SpO2 measurements. The Docsun Telehealth Portal predicted systolic BP with an accuracy of 94.81% and diastolic BP with an accuracy of 95.71%. Conclusions The results of the study show that the accuracy of the HR, BP, SpO2, and RR values raised by the Docsun Telehealth Portal, compared against the clinically approved medical devices, proved to be accurate by meeting predefined accuracy guidelines.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.003
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.105
Threshold uncertainty score0.577

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0030.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.001
Science and technology studies0.0000.000
Scholarly communication0.0000.001
Open science0.0000.000
Research integrity0.0000.001
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.037
GPT teacher head0.365
Teacher spread0.328 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations3
Published2024
Admission routes1
Has abstractyes

Explore more

Same venueJMIR Formative ResearchSame topicNon-Invasive Vital Sign MonitoringFrench-language works237,207