Record W4406690298 · doi:10.1016/j.jmbbm.2025.106907

Effect of head size, head material, and radiation dose on the repeatability of CT-RSA measurements of femoral head penetration

2025· article· en· W4406690298 on OpenAlexafffund
J Polus, Bart L. Kaptein, Brent A. Lanting, Matthew G. Teeter

Bibliographic record

VenueJournal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · 2025
Typearticle
Languageen
FieldMedicine
TopicOrthopaedic implants and arthroplasty
Canadian institutionsWestern University
FundersNatural Sciences and Engineering Research Council of CanadaArthritis Society
KeywordsRepeatabilityHead (geology)Femoral headMaterials scienceNuclear medicinePenetration (warfare)Biomedical engineeringMedicineAnatomyMathematicsGeologyStatistics

Abstract

fetched live from OpenAlex

BACKGROUND: The risk of early revision of total hip arthroplasty (THA) for polyethylene wear is now low, but there remains a need to perform wear measurements in patients for clinical surveillance. The gold standard of wear measurements has been radiostereometric analysis (RSA), which has limited availability. The use of computed tomography (CT) to perform THA wear measurement was described a decade ago and found to have acceptable accuracy and precision, but high radiation dose was a concern. Additionally, the use of larger femoral head sizes and ceramic femoral heads has risen in the past decade. The objectives of the study were to determine the effect of femoral head size, femoral head material, and lowered radiation dose on femoral head penetration measurement repeatability. METHODS: A cadaveric hip was implanted with a cementless THA implant system. CT scans were acquired at a conventional radiation dose and at a reduced dose and repeated for a 32 mm and 36 mm cobalt-chromium femoral heads and a 32 mm ceramic femoral head. Apparent translation of the femoral head versus the acetabular cup was measured between the repeated scans using a CT-RSA software, where deviations from zero indicated measurement precision. RESULTS: The mean and standard deviation of translations in all planes was <0.200 mm. There was no effect for 3D translation of increasing cobalt-chromium head size (p = 0.2252). Cobalt-chromium heads had superior repeatability compared to ceramic heads at reduced dose (p = 0.022), but not at conventional dose (p = 0.1265). Further, superior repeatability was achieved with the reduced dose scan for the cobalt-chromium head (p = 0.0058), however there was no difference between doses for the ceramic head (p = 0.8148). DISCUSSION: CT-based wear measurement repeatability is excellent and consistent with prior literature even when implementing a larger femoral head, a ceramic femoral head, or reducing radiation dose to 25% of a conventional clinical scan.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.005
metaresearch head score (Gemma)0.018
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.005
Threshold uncertainty score0.026

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0050.018
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.001
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.029
GPT teacher head0.326
Teacher spread0.297 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations3
Published2025
Admission routes2
Has abstractyes

Explore more

Same venueJournal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials→Same topicOrthopaedic implants and arthroplasty→French-language works237,207→