Validating the accuracy of Omron J760 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020

2025· article· en· W4407885599 on OpenAlexaff
Bihua Chen, Guorong Chen, Hong Jiang, Ya Sun, Jie Chen, Dan Wang, Mingzhi Chen, Jia Shi, Yan Peng, Yimin Cheng, Yang Cheng, Yuanyuan Ding, Jin Su, Ming Liu, Fulai Shen, Yi Shen, Qiyun Cao, Jingzhi Su

Bibliographic record

VenueBlood Pressure · 2025
Typearticle
Languageen
FieldMedicine
TopicBlood Pressure and Hypertension Studies
Canadian institutionsArtificial Intelligence in Medicine (Canada)
Fundersnot available
KeywordsMedicineBlood pressurePopulationInternal medicineEnvironmental health

Abstract

fetched live from OpenAlex

PURPOSE: The accuracy of the Omron J760 electronic blood pressure (BP) monitor for upper arm BP measurement in the adult general population was validated following the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 protocol. We expect that this device can be used for home blood pressure monitoring, with its measurements serving as a clinical reference. METHODS: Subjects meeting the age, gender, BP, and cuff distribution criteria as specified by the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 were recruited. BP measurements were conducted using the same-arm sequential method. The test device cuff was suitable for arm circumferences ranging from 22.0 cm to 42.0 cm. RESULTS: Eighty-nine participants were initially recruited, 85 were evaluated after excluding 4 participants. The mean age of the participants was 48.5 ± 15.17 years. For validation criterion 1, the mean ± standard deviation (SD) of the differences between the test device and the reference BP measurements were 0.2 ± 5.74 mmHg for systolic BP and -0.9 ± 4.69 mmHg for diastolic BP. These results met the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd1:2020 standard, which requires differences of ≤5 ± ≤8 mmHg. For validation criterion 2, the mean differences between the test device and the reference device were 0.2 ± 5.10 mmHg for systolic BP and -0.9 ± 4.30 mmHg for diastolic BP. This criterion was satisfied with criterion 2 of ≤6.95 mmHg for systolic BP and ≤6.88 mmHg for diastolic BP. CONCLUSION: The Omron J760 electronic BP monitor meets the requirements of the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd 1:2020 validation standard.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.023
metaresearch head score (Gemma)0.038
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.023
Threshold uncertainty score0.121

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0230.038
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0010.001
Science and technology studies0.0010.001
Scholarly communication0.0010.001
Open science0.0020.001
Research integrity0.0020.001
Insufficient payload (model declined to judge)0.0010.002

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.020
GPT teacher head0.300
Teacher spread0.280 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations1
Published2025
Admission routes1
Has abstractyes

Explore more

Same venueBlood PressureSame topicBlood Pressure and Hypertension StudiesFrench-language works237,207