WE-AB-303-10: The Use of On-Board KV Imaging During Respiratory-Gated VMAT Delivery to Determine the Correlation and Phase Shift Between External Marker Motion and Internal Tumour Motion

2015· article· en· W949339994 on OpenAlexaff
Ilma Xhaferllari, Omar El‐Sherif, Stewart Gaede

Bibliographic record

VenueMedical Physics · 2015
Typearticle
Languageen
FieldMedicine
TopicAdvanced MRI Techniques and Applications
Canadian institutionsWestern University
Fundersnot available
KeywordsImaging phantomTruebeamGatingBreathingCorrelation coefficientPhysicsMedical imagingNuclear medicineBiomedical engineeringCorrelationRespiratory monitoringRespiratory systemLinear particle acceleratorOpticsMedicineMathematicsRadiologyStatisticsAnatomy

Abstract

fetched live from OpenAlex

Purpose: To determine the correlation between internal tumor motion and external surrogate motion and any potential phase shifts during respiratory-gated VMAT delivery using on-board kV imaging. Methods: A respiratory motion phantom was used to trace the motion of an internal target, embedded with a 3 cm moving target, and an external surrogate. On-board kV projections were acquired at 11 frames per second during VMAT delivery on a TrueBeam v.1.6 (Varian Medical Systems, Palo Alto, CA) operating in developer mode. Full motion encompassed treatment delivery using sinusoidal motion with 4 second period and 2 cm peak-to-peak amplitude, and real-patient breathing (RPB) motion was programmed to the phantom. Respiratory-gated treatment deliveries (30–60% gating window) on five additional RPB waveforms were programmed to investigate correlation within the gating window. To investigate the capability of during treatment kV imaging to detect phase shifts during gated delivery, a second phantom with independent motion was used. Eight controlled phase shifts in intervals of 0.4 seconds were added to the sinusoidal motion. For each phase shift, the Pearson linear correlation coefficient statistical test was performed for each set of respiratory traces. The phase shift was calculated by shifting the external trace, in time, until maximum correlation was obtained. Results: A strong internal and external correlation was obtained for both the free-breathing (sinusoidal R2=0.993 and RPB R2=0.990) and respiratory-gated cases (range from R2=0.986–0.996). There was no significant difference between the programmed shift and that acquired using on-board kV imaging (p=0.899, R-squared = 0.997). Conclusion: During treatment kV imaging has the capability to verify intrafractional anatomical position. It can be accurately used as a tool to quantify internal and external correlation, and determine phase shifts within the gating window, and ultimately, the accuracy of respiratory-gating treatment delivery.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.001
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.002
Threshold uncertainty score0.007

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0000.001
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0020.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.059
GPT teacher head0.330
Teacher spread0.270 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2015
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced MRI Techniques and Applications→French-language works237,207→